Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
ANTITHROMBOGENISITY OF TUBU COATED BY THE HYDROGEL WITH LONG POLYETHYLENEOXIDE CHAIN (PEO)
-CLINICAL STUDY AS THE TUBE FOR DRAINAGE AFTER CARDIAC SURGERY
Y. OGURAH. NAKAJIMAJ. ANBEM. OZEKIT. MITSUISHIR. SUZUKIT. AKASAKAO. KAWANAMIY. FUTAMIN. NOGUCHI
Author information
JOURNAL FREE ACCESS

1990 Volume 19 Issue 3 Pages 1083-1086

Details
Abstract

Antithrombogenicity of the tube coated by PEO (PEOtube) was compared with polyvinylchloride tube (PVCtube), and clinicaly evaluated as the tube for drainage after cardiac surgery. By the study of scanning electron micrograph, fibrin formation on the surface of the PEOtube was inhibited comparing with the PVCtube. Patency of the PEOtube was better than PVCtube significantly (p<0.01) . Therefore it is suggested that longer drainage is possible by the PEOtube.

Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback
Top