CLINICAL EXPERIENCE OF LONG TERM USE OF A NEWLY DEVELOPED MEMBRANEOXYGENATOR (MENOX, AL-2000)
A NAKASHIMAR TOMINAGAF FUKUMURAY TSURUHARAM HISAHARAY KANEGAEK KURISUK SHIRAISHIH YASUIK. TOKUNAGA
Author information
JOURNAL FREE ACCESS

1992 Volume 21 Issue 2 Pages 688-691

Details
Abstract
We clinically evaluated the gas transfer capacity and the durability of a newly developed membrane oxygenator “MENOX, AL-2000”. The oxygenator was set in a circuit for veno-arterial bypass in 3 patients. No serum leakage was seen and no significant deterioration of gas transfer performance was observed up to 211 hours. Moderate amount of fibrin clot was observed mainly at the inlet portion although activated clotting time was maintained at approximately 200 seconds under anticoagulation therapy using with heparin and nafamostat. In conclusion, MENOX, AL-2000 is a suitable oxygenator for long-term VA bypass support.