Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
CONSIDERATION OF THE HARDWARE TO DEVELOP NEW ARTIFICIAL OXYGENATORS
T. SUEDAH. SHINJIS. FUKUNAGAY. MATSUURA
Author information
JOURNAL FREE ACCESS

1997 Volume 26 Issue 3 Pages 811-814

Details
Abstract
We have devised the following artififial lungs and evaluated the hardware of these devices. 1) An outside flow membrane oxygenator using a silicone hollow fiber was developed and is available for clinical use. This oxygenator is useful in selective cerebral perfusion. Modification such as heparin coating and minimization of the shape is desirable. 2) Liquid-liquid oxygenators using oxygenated perfluorocarbon (PFC) were devised. These oxygenator had an ability of oxygenation without membrane oxygenator but demerit of thromboembolism of PFC. 3) Intravascular balloon oxygenator (IVBO) was devised and evaluated experimentally. An animal experiment in which the IVBO was implanted into the inferior vena cava of a mongrel dog, showed an increase of venous return and oxygen pressure of the venous blood during pumping. Futher modification is hopeful in progress of the membrane material.
Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback
Top