Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
Micropump for artificial metabolic and endocrine organs
N. KABEIM. KILUCHIA. YAMADAE. MACHIYAMAY. SAKURAI
Author information
JOURNAL FREE ACCESS

1978 Volume 7 Issue 1 Pages 105-108

Details
Abstract
The roller type micropump driven by the pulse motor was developed as the insulin infusion pump for artificial cell. Flow rate of this pump is able to control by output signals of microcomputer. The volume displacement for one pulse was about 0.006μl. The pump output was increased in quite linear fashion from 0.003 to 0.1ml/min as increase the pulse number. The reproducibility was better than 2 % and pump flow rate was independent of load pressure.
As the next version of this type of pump, smaller one was also developed. Dimensions and weight of this pump are almost same as those of an usual wrist watch, so that it can be put on the patient's hand.
Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback
Top