体内Compartmentモデルを用いた透析システムの化学工学的考察
酒井 清孝坂口 保雄鈴木 満東 仲宣品川 茂
著者情報
ジャーナル フリー

1978 年 7 巻 2 号 p. 322-325

詳細
抄録
In this study a new parameter CL/B. W. was considered in analysing the artificial kidney system (A. K. S.). Using this parameter, the transfer phemomena of thre three kinds of solute in A. K. S. was examined by 2-pool and new model.
Those were urea, a solute stored in tissue like creatinine and a solute combined with some kind of protein.
The urea transfer property under any values of clearance in A. K. S. became equal if CL/B. W. was decided to be constant. The transfer amount of the solute that has a property of storage in tissue or protein binding was decreased inversely propotional to the quantity of storage in tissue or binding to prtein.
著者関連情報
© 一般社団法人 日本人工臓器学会