人工血液の体外循環への応用
安達 盛次川島 康生森 透橋本 聡一高野 久輝広瀬 一大山 朝賢横田 博雅中田 精三浜路 政靖村田 弘隆奥田 彰洋金 勢華藤田 毅曲直部 寿夫
著者情報
ジャーナル フリー

1978 年 7 巻 2 号 p. 422-425

詳細
抄録
Extracorporeal circulation was performed in dogs with a perfusate containing 0.6 to 2.5Gm/dl of hemoglobin and 11 to 20Gm/dl of Perfluorochemical emulsion (PFC). The oxygen consumption increased as the oxygen delivery was enhanced by increasing the perfusion rate. The consumption however, did not increase over a certain value even though the delivery was raised over 8ml/Kg B. W./min. The maximal oxygen consumption was a_little less than 5ml/Kg B. W./min at 37°C, and this amount of oxygen consumpton was obtained when the oxygen delivery increased over 8ml/Kg B. W./min. The minimal perfusion rate necessary to meet the maximal oxygen consumption in dogs was about 120ml/Kg B. W./min when the arterial oxygen content was kept in the range of 6 to 7 volume % by the use of PFC, even though the hemoglobin content was reduced to less than 1.2Gm/dl.