人工臓器
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
閉鎖回路をもってする体外循環の方法 第2報
山田 公弥日笠 頼則松田 光彦大頭 信義青嶋 実村田 真司村田 雄彦龍田 憲和相馬 敬司西村 営三
著者情報
ジャーナル フリー

1978 年 7 巻 2 号 p. 455-458

詳細
抄録
Total heart-lung bypass wit h closed circuit was performed under automatically controlled CVP, and oxygen consumption and fraction rate of SVC and IVC return were studied in our previous report.
In this report histological changes of the liver, portal vein pressure, cerebral drainage and total systemic peripheral resistance during heartlung bypass were discussed in variously controlled CVP. Consequently, during our closed circuit heart-lung bypass, 7-9 cmH2O CVP was recommended to minimize histological and metabolic changes and to keep venous return stable.