Björk-Shiley弁による大動脈弁置換術の成績
―特に弁機能について―
河内 寛治森 透北井 惣一郎中埜 粛大山 朝賢井原 勝彦森崎 弘士島崎 靖久八木原 俊克榊原 哲夫川島 康生横田 侃児
著者情報
ジャーナル フリー

1979 年 8 巻 1 号 p. 132-135

詳細
抄録

We evaluated late the in vivo hemodynamic function of Björk-Shiley prostheses in 15 patients who had undergone aortic valve replacement. Cardiac catheterization was performed at rest and during exercise in 5 pts with size 19·21·23 valves (Group I) in an average of 45 months, and in 10 pts with size 25·27·29 valves (Group II) in an average of 10 months after operation.
The left ventricle was catheterized by the transseptal technique postoperatively.
The average aortic mean systolic gradient in G-I was 25mmHg at rest and 46mmHg during exercise. It in G-II was 5mmHg at rest and 6mmHg during exercise.
The average calculated orifice area in G-I was 1.14cm2 at rest and 1.27cm2 during exercise. And it in G-II was more than 2.0cm2.
Left ventricular function such as cardiac index, end-diastolic volume and ejection fraction was unchanged for the group. However, LVEDP in 2 pts in G-I elevated abnormally at rest. Thus, patients in G-I had functional