D50 金属ならびにプラスチック製人工骨頭の臼蓋軟骨に対する影響
中島 育昌伊丹 康人赤松 功也富田 泰次永井 素大唐沢 重雄神前 智一天神 宏純加藤 善久
著者情報
ジャーナル フリー

1979 年 8 巻 1 号 p. 238-241

詳細
抄録

Femoral heads in seven dogs were replaced with metal (titanium) and high density polyethylene (HDP). Clinical and roentgenographic evaluations and histopathological, scan electron microscopic and metal microscopic examinations were done in every dogs at intervals from 3 months to 3 years after surgery.
As a conclusion, HDP femoral head was inadequate for use as the artificial femoral head, because degeneration of cartilage or wear and deformity of HDP femoral head were occured 3 months after surgery and such destruction changes developed until 3 years after surgery.
On the other hand, titanium femoral head as metal was less harmful for acetabular cartilage even if 3 years after surgery, if it was not so larger. Then an application of metal femoral head for clinical use is recommended.