腎保存装置の開発に関する研究
小崎 正已東 仲宣春日 幸夫佐々木 裕一玉置 勲盧 建基鈴木 満工藤 武彦辻 孝彦稲村 征夫坂井 直司高橋 雅俊
著者情報
ジャーナル フリー

1980 年 9 巻 2 号 p. 351-354

詳細
抄録
Continuous hypothermic perfusion is today an accepted method of renal preservation in the world. Some preservation machine such as Belzer, Waters (Mox 100), Travenol and Gambro are clinically used in U. S. A. and Europe. However, perfusion method of kidney preservation is not yet clinically used in this country. We have been developing a new preserving machine that would be easier to handle and cheaper to purchase. This machine is consisted of an organ chamber, two roller pumps, a membrane oxygenator, a heart exchanger, a refrigerator, two in-line thermistor probes and two automatic pressure controllers. An organ chamber has double walls which is consisted of inside and outside walls. Cold water flows between two walls to cool an organ chamber. A pressure controller maintains arterial systolic pressure to the indicated one, which is usually 60mmHg, and when the pressure rises, the controller decreases the rotation of the roller pump, making the renal flow rate lower. Belzer's cryoprecipitated plasma was used as the perfusate and we succeeded in preserving canine kidneys using this machine for 72-96 hrs.
著者関連情報
© 一般社団法人 日本人工臓器学会