Journal of Advanced Science
Online ISSN : 1881-3917
Print ISSN : 0915-5651
ISSN-L : 0915-5651
Letter
機能性高分子を用いたアルコール計測用フレキシブルセンサ
KAJIRO TakumiSHIRAI TakayukiYANO KazuyoshiHIRANUMA YoshitakaHIRAMATSU HideoCHU Ming XingTAKAHASHI DaishiARAKAWA TakahiroKUDO HiroyukiMITSUBAYASHI Kohji
著者情報
ジャーナル フリー

2010 年 22 巻 3+4 号 p. 36-37

詳細
抄録

A flexible alcohol sensor using functional polymers for transcutaneous bioinstrumentation was constructed and tested. The sensor was composed of a flexible electrode and an enzyme membrane. In the fabrication process of the electrode, Pt and Ag/AgCl electrodes were formed onto a PDMS membrane by MEMS technique. In the enzyme immobilization, a mixture solution of alcohol oxidase and photocrosslinkable polyvinyl alcohol was spread onto a PTFE membrane. The AOD was immobilized by ultraviolet irradiation. The enzyme membrane was attached to the sensing region of the electrode with biocompatible polymer. In the characteristic evaluations, the sensor was dipped into a phosphate buffer, and then a potential of +400 mV (vs. Ag/AgCl electrode) was applied to Pt-working electrode. The change of the output current by dropping of ethanol solutions was measured with PC. As a result of the evaluation, the output current rapidly increased immediately after the dropping of alcohol solutions and then was stable within 60 seconds. The output current linearly corresponded to alcohol concentrations over a range of 0.001-7.0 mmol/l, with a correlation coefficient of 0.999.

著者関連情報
© 2010 Society of Advanced Science