2020 Volume 11 Issue 2 Pages 138-142
【Background】Prismaflex®, which was launched in 2018, this feature is expected to reduce coagulation within the circuit. 【Method】We divided TR-55X® (Toray Medical) and Prismaflex® (Baxter) used in our hospital, in order to retrospectively compare the effects on coagulation within the circuit. 【Results】Frequency of unexpected circuit exchange was significantly higher in the P group (p=0.003). The returning blood pressure was significantly higher in the TR group (p=0.003). There were four “other causes” other than increased return blood pressure and increased inlet pressure only in the P group. 【Conclusion】There were many unexpected circuit changes with P group, due to blood coagulation in the hemofilter or deviations in the operating procedure; consequently, going forward, it was considered that the results could be improved by improving the hemofilter and mastering the operation.