Three-Dimensional Quantification of Cardiac Surface Motion
A Newly Developed 3-Dimensional Digital Motion-Capture and Reconstruction System for Beating Heart Surgery
Toshiki WatanabeMotoki OdamuraMasahumi OkadaYoshihiko NakamuraSadao OmataHitoshi Yokoyama
Author information
JOURNAL FREE ACCESS

2007 Volume 8 Issue 4 Pages 409-415

Details
Abstract
OBJECTIVE: This study was aimed to characterize quantitatively the three regional cardiac surface motions before and after stabilization using a stabilizer (Octopus 4.3, Medtronics Inc.) by our newly developed “3-dimensional digital motion-capture and reconstruction system” in an animal experiment setting. METHODS: Ten pigs underwent full sternotomy. Three tiny metallic markers (diameter 2 mm) coated with a reflective material were attached on three regional cardiac surfaces (LAD, RCA and LCX). These makers were captured by two high-speed digital video cameras (955 frames/s) as 2-dimensional data, and reconstructed to 3-dimensional data points. RESULTS: Virtual 3-dimensional images of the beating heart surface were successfully reconstructed and details of the three regional cardiac surfaces were obtained. CONCLUSIONS: This system is useful for a precise 3-D quantification of the heart surface movement, and development of a better stabilizer for beating heart surgery.