Determination of Pranoprofen in Serum by Automated Column-Switching High-Performance Liquid Chromatography
疋田 興造井上 義政小島 章生大倉 洋甫
著者情報
ジャーナル フリー

1989 年 18 巻 4 号 p. 180-183

詳細
抄録
An automated method for the determination of pranoprofen, an anti-inflammatory drug, in human serum has been developed by means of column-switching high-performance liquid chromatography with ultraviolet detection (247 nm).A bovine serum albumin-coated ODS column (MK precolumn BSA-ODS) and a reversed phase column (TSK gel ODS-80TM) were used as the sample preparation column and analytical column, respectively. A twotimes diluted serum sample (20 μl) was applied onto the sample preparation column. The detection limit of pranoprofen is 0.05?Eg/ml serum. The recovery of pranoprofen added to serum (2.0-8.0 μg/ml) was approximately 100% with a standard deviation of 3% or less.