臨床化学
Online ISSN : 2187-4077
Print ISSN : 0370-5633
ISSN-L : 0370-5633
DEAE-Sephadex A-50を用いた小カラム法の臨床検査への応用 (第2報) ヒト血清におけるGOTmの定量
石黒 伊三雄長村 洋一荻津 直通大槻 隆明篠原 力雄
著者情報
ジャーナル フリー

1977 年 5 巻 2 号 p. 163-169

詳細
抄録
A new mini-column method, which we devised previously for determination of IgG, was used for quantitative determination of the enzyme activities of GOTm in human serum. When 50μl of serum was applied to this mini-column with DEAE-Sephadex A-50, GOTm was not adosorbed and eluted rapidly. The caliblation curve of purified human GOTm was determined with Karmen's, Babson's and Reitman-Frankel method, and Karmen's method was the best one of these method. When several sera with addition of purified GOTm was applied to this column, GOTm was recovered about 100% in eluate. The correlation of this method and electrophoresis method was well at low activity of GOTm, but not well at high activity of GOTm by the reason why electrophoresis method can not determined GOTm accuratly. The correlation of this method and immunoadsorbance method was very well. This method thus appeares to be useful in daily clinical test.
著者関連情報
© 日本臨床化学会
前の記事 次の記事
feedback
Top