炎症
Online ISSN : 1884-4006
Print ISSN : 0389-4290
ISSN-L : 0389-4290
ルミノール依存性化学発光測定による迅速簡便な白血球機能検査の基礎的条件
小橋 修小橋 悠紀子豊島 静枝重松 信昭
著者情報
ジャーナル フリー

1985 年 5 巻 3 号 p. 205-210

詳細
抄録

A simple and rapid microtechnique using 10 ul of whole blood was developed for studying a luminol-dependent chemiluminescence (CL) response. This assay have been thought to be unreliable because of poor reproducibility. In fact, the CL response of this assay was intrinsically affected by contamination of red blood cells or other cellular components and plasma constitutents. More importantly, there were several serious problems to get a reliable and reproducible measurement, because this CL respons by itself was varied hours by hours and day by day and different machines gave us a significantly different measurement to the CL response of the same blood sample. We carefully studied the effect of red blood cells and serum on CL response and found that if 10 ul of whole blood were used instead of 100 ul, this assay can be successfully and reliably used for evaluating the phagocytic function of peripheral blood phagocytes, especially by a simultaneous measurement of CL respons against PMA and OZ immediately after bleeding at a certain fixed time of a patient.
This simultaneous microtechnique of whole blood CL response will offer a very important knowledge of cell surface functions (FcR and C3bR) of phagocytic cells, involved in the pathological events of various diseases.

著者関連情報
© 日本炎症・再生医学会
前の記事 次の記事
feedback
Top