日本リウマチ・関節外科学会雑誌
Online ISSN : 1884-9059
Print ISSN : 0287-3214
ISSN-L : 0287-3214
RAにおけるDOH式人工肘関節置換術
―3年以上例の検討―
近藤 泰紘八野田 実上田 俊一新田 雅英中尾 浩志
著者情報
ジャーナル フリー

1992 年 11 巻 2 号 p. 197-204

詳細
抄録
The DOH elbow prosthesis is a non-constrained type composed of a humeral mold with a stem (made of COP from 1985 and titanium alloy from 1987) and an HDP ulnar component with a stem.
The clinical and radiological results of 48 total elbow replacements in which such a prosthesis was implanted in 35 patients with rheumatoid arthritis were reviewed retrospectively. These cases were followed over at least 3 years (average 4.3 years) .
Final average motion did not decrease as compared with 3 months after the operation. Extension and supination were rather improved at the time of follow-up, and on radiographs there were no remarkable changes around ulnal components, but radiolucent lines were observed around humeral components in 18 elbows. The incidence of proximal migration was very high in non-cemented humeral components (5 in 9 elbows), and low in cemented humeral components (6 in 3 9 elbows) . Progression of the migration had a tendency to stop in time.
著者関連情報
© 日本リウマチ・関節外科学会(2006年~:2005年以前は投稿規程に著作権に関する記載なし)
前の記事
feedback