The Journal of Japan Society for Laser Surgery and Medicine
Online ISSN : 1881-1639
Print ISSN : 0288-6200
ISSN-L : 0288-6200
Safeguards against CO2 Laser Irradiation in Dentistry
Kouji NISHIKAWAAkinori NAGASAWAKazuichi KATOYoshiaki HIRATSUKATsuneo KAWAHARAKazuhiko ATSUMI
Author information
JOURNAL FREE ACCESS

1980 Volume 1 Issue 1 Pages 287-293

Details
Abstract

This paper reports on dental laser safety investigation and CO2 laser protectors for dentistry
1) The laser protect effects of the experimental materials. 2) Reflection and heat product ion on the surface of metal instruments after CO2 laser irradiation. 3) Development of prototype of CO2 laser protectors for dental use. This protector i s composed of “WASHI”(Japanese paper) filled with water and aluminum foil.
Results
1) Both “WASHI” contained water and aluminum foil showed excel lent laser inhibition effects. 2) Black coated metal surface with various materials was effective to prevent reflection of CO2 laser, while the temperature on it was elevated. 3) The prototype laser protectors were able to inhibit soft and hard tissues from CO2 laser energy, even in 30 watts for 2 seconds.

Content from these authors
© Japan Society for Laser Surgery and Medicine
Previous article Next article
feedback
Top