経頭蓋磁気刺激における脳と刺激用コイルの高精度位置計測システムの開発
井上 朋紀, 三島 行雄, 樋脇 治
著者情報
ジャーナル フリー

2006 年 44 巻 4 号 p. 674-679

詳細
抄録
Transcranial magnetic stimulation (TMS) is a method for noninvasive stimulation of cerebral cortex, and it has contributed to clinical and basic researches of brain function. In order to estimate the accurate stimulating points of the cortex in TMS, precise measurement of the subject's head and the stimulating coil is necessary. In this study, we have developed the positioning TMS system with a three-dimensional (3-D) digitizer and a multi-articular system. We proposed a method for the accurate measurement of a subject's head and cortex, in which the location data of the subject's face surface captured by a 3-D digitizer were superimposed on the magnetic resonance imaging (MRI) data of the subject's face surface. Using this system, the precise estimation of the stimulated sites of the cortex in TMS was achieved. The validity of the system was verified by the experiment on the TMS of the motor cortex.