画像誘導下手術支援ロボットシステムのためのMRI対応モータ駆動法
鈴木 孝司廖 洪恩小林 英津子佐久間 一郎
著者情報
ジャーナル フリー

2006 年 44 巻 4 号 p. 728-734

詳細
抄録

For robotic surgery assistance systems that use magnetic resonance imaging (MRI) for guidance, the problem of electromagnetic interference is common. Image quality is particularly degraded if motors are running during scanning. We propose a novel MRI-compatible method considering the pulse sequence of imaging. Motors are driven for a short time when the MRI system stops signal acquisition (i.e., awaiting relaxation of the proton), so the image does not contain noise from the actuators. The MRI system and motor are synchronized using a radio frequency pulse signal (8.5 MHz) as the trigger, which is acquired via a special antenna mounted near the scanner. This method can be widely applied because it only receives part of the scanning signal and neither hardware nor software of the MRI system needs to be changed. As a feasibility evaluation test, we compared the images and signal-to-noise ratios between the cases with and without this method, under the condition that a piezoelectric motor was driven during scanning as a noise source, which was generally used as a MRI-compatible actuator. The results showed no deterioration in image quality and the benefit of the new method even though the choice of available scanning sequences is limited.