MRA画像における脳血管名の自動対応付け手法の開発
山内 将史内山 良一小椋 潤横山 龍二郎原 武史安藤 弘道山川 弘保岩間 亨星 博昭藤田 広志
著者情報
ジャーナル フリー

2007 年 45 巻 1 号 p. 27-35

詳細
抄録

The detection of unruptured aneurysms is a major task in magnetic resonance angiography (MRA). However, it is difficult for radiologists and/or neurosurgeons to detect small aneurysms on maximum intensity projection (MIP) images because adjacent vessels may overlap with the aneurysms. Therefore, we proposed a method for making a new MIP image, the SelMIP image, containing interested vessels only by manually selecting a cerebral artery from a list of cerebral arteries recognized automatically. For the automated classification of cerebral arteries, two three-dimesional images, a target image and a reference image, were compared. Image registration was performed using global matching and rigidity transformation. The segmented vessel regions were classified into eight cerebral arteries by calculating the Euclidian distance between a voxel in the target image and each of the voxels in the eight labeled vessel regions in the reference image. In applying the automated cerebral arteries recognition algorithm to 110 MRA studies, the results of subjective evaluation were that 76.4% (84/110) were rated as good, 13.6% (15/110) as adequate, and 10.0% (11/110) as poor. The results rated good or adequate are considered acceptable and would be adequate for clinical use. Overall, 90.0% (99/110) of MRA studies attained a clinically acceptable result. Our new viewing technique will be useful in assisting radiologists to detectaneurysms and reducing the interpretation time.