生体医工学
Online ISSN : 1881-4379
Print ISSN : 1347-443X
ISSN-L : 1347-443X
研究
有限要素法解析に基づく経皮的PA信号検出器の高感度化に関する検討
和田森 直石原 康利
著者情報
ジャーナル フリー

2008 年 46 巻 2 号 p. 238-245

詳細
抄録
In order to measure noninvasively living body information using photoacoustic spectroscopy(PAS), fundamental investigation regarding a high sensitivity detector was carried out by the finite element method(FEM). The mathematical model reflecting the physical mode regarding a signal generation, propagation and detection of PA signal was used, and the analysis for optimizing PA detector cell was carried out. First, in order to detect the PA signal with high sensitivity, it was shown that it is necessary to make total volume of a cell small. Next, the influence on sensitivity due to in the gap between a target sample and a detector cell was evaluated. As a result, it was shown that there is not a large influence on the detection of the PA signal, even if both degree of adhesion changes slightly. Furthermore, the sensitivity of a resonant type cell and a non-resonant type cell was compared, and it was shown that the sensitivity of the resonant type is about 9-120 times that of the non-resonant type. From these results, the conditions for achieving the PA detector cell with high sensitivity were suggested.
著者関連情報
© 2008 社団法人日本生体医工学会
前の記事 次の記事
feedback