生体医工学
Online ISSN : 1881-4379
Print ISSN : 1347-443X
ISSN-L : 1347-443X
研究
MRIによる術具の位置・姿勢計測法に関する研究
小野木 真哉廖 洪恩小林 英津子佐久間 一郎
著者情報
ジャーナル フリー

2009 年 47 巻 6 号 p. 582-587

詳細
抄録
The combination of MR guided surgery and computer aided surgery has a potential to ensure accurate, certain, and safe procedures. Although a tracking device is indispensable for computer assisted surgery, conventional tracking devices are not available or difficult to use in a MR scanner because of occlusion and strong magnetic field. We proposed a novel tracking method by using MR scanner. The method provides three dimensional positions and orientations of a miniature receiver coil (tracking coil) consisted of three marker coils and tuning capacitor, and matching capacitor. We describe experiments we conducted to evaluate the basic performance of the method. First, a surgical tool affixed the tracking coil was calibrated. Next, measurement accuracy of the tool's tip position and orientation was evaluated. The error of root mean square was 0.92 mm and 0.47 degree. This paper recommends a tracking method for computer assisted surgery with MRI to enable tracking in MR environment. Results obtained with the method demonstrate its feasibility.
著者関連情報
© 2009 社団法人日本生体医工学会
前の記事 次の記事
feedback