Abstract
Pulse oximetry is used world wide in medical practice. Development of theory of pulse oximetry is still necessary for improving its performance. First we hypothesized a completely scattered light field in the object, and adopted Sir Arthur Schuster's scattering theory as the basis of theory of pulse oximetry. Second, a theoretical formula for pulse oximetry was proposed considering the arterial blood and tissue. Third, with a volunteer experiment the theoretical formula was proved and some constants were obtained.