A new proposal for setting servo-control reference for accurate measurement of instantaneous arterial pressure based on the volume-compensation method
Masanori KakimotoRyouta SuzukiJihyoung LeeKenta MatsumuraTakehiro YamakoshiShinobu TanakaMasamichi NogawaKen-ichi Yamakoshi
Author information
JOURNAL FREE ACCESS

2014 Volume 52 Issue Supplement Pages O-520-O-521

Details
Abstract
Servo-control reference corresponding to arterial unloading volume (r_V0) is one of the most important parameters when measuring accurately blood pressure (BP) using the volume-compensation method (VCM). We proposed here a new method to set r_V0 by adjusting it to yield approximately equalized values in systolic and diastolic BP that were determined by a pre-operation procedure of the tube-law based volume-oscillometric method. BP measurements at a finger were made in 10 volunteers by the VCM with r_V0 settings of both the original (r_V0_org) and the new value (r_V0_prp), while measuring simultaneously brachial BP as a reference. Analyses revealed that r_V0_prp was higher than r_V0_org, and that the correlation coefficients between systolic/diastolic BP values using the r_V0_prp and the r_V0_org settings and those by the brachial BP measurements were 0.97 and 0.83, respectively. These results indicate the effectiveness of this setting method for more practical and accurate BP measurement by the VCM.