生体医工学
Online ISSN : 1881-4379
Print ISSN : 1347-443X
ISSN-L : 1347-443X
一般演題(口演)
Magnetocaloric property of mechanically fabricated feromagnetic implants for magnetic hyperthermia
Kenji KodamaTatsumi IchiseYukinori TaniguchiManabu FukugauchiYasuhiro MikiMakoto AsanoNobuyoshi HosoitoHiroki Kuniyasu
著者情報
ジャーナル フリー

2014 年 52 巻 Supplement 号 p. O-94-O-95

詳細
抄録
The essential requirements for implants utilized in magnetic hyperthermia are high calorific power in alternating magnetic field, high corrosion resistance in body. In addition, the fabrication process is require simple and stable with low cost from a practical application perspective. In this research, we investigate calorific property of micro implants dispersed ferromagnetic particles. The implants was molded into beads needle and capsule by simple mechanical processing. The amount of heat generation of ferromagnetic particles was increased with increasing magnetic hysteresis loss by mechanical milling treatment which is able to pulverize and provide strong strain. Temperature rising of the implants in chicken meat as a mock tumor and thermal alteration of surrounding cells will be reported.
著者関連情報
© 2014 Japanese Society for Medical and Biological Engineering
前の記事 次の記事
feedback