生体医工学
Online ISSN : 1881-4379
Print ISSN : 1347-443X
ISSN-L : 1347-443X
オーガナイズドセッション
大動脈基部装着軸流型補助人工心臓の基礎開発と現状
岡本 英治矢野 哲也石田 裕也三田村 好矩
著者情報
ジャーナル フリー

2014 年 52 巻 Supplement 号 p. OS-7-OS-8

詳細
抄録
We have developed a miniature axial flow pump Valvo Pump 2 that is intend to install the base of the ascending aorta. The new axial flow pump consists of a stator, an impeller in which four neodymium magnets are incorporated, and the impeller is directly driven by the stator. Rotation of the impeller is suspended by a passive magnet bearing in an axial direction, and by a hydrodynamic bearing in a radial direction. The magnetic passive bearing tolerates an axial load of 3.5 N, and the CFD analysis shows that the magnetic passive bearing can sustain the impeller in an axial direction. The hydrodynamic bearing is currently under development to obtain enough radial bearing capacity.Further studies are required, the new axial flow pump Valvo Pump 2 could be promising to be a new VAD.
著者関連情報
© 2014 社団法人日本生体医工学会
前の記事 次の記事
feedback