Prediction of fatigue failure of coronary stent under cyclic bend loading using finite element analysis
KIYOTAKA IWASAKIXiaodong ZHUHirotsugu TOKUTAKEMitsuo UMEZU
Author information
JOURNAL FREE ACCESS

2015 Volume 53 Issue Supplement Pages S136_01

Details
Abstract
Introduction: Coronary stent fracture is recognized as causes of restenosis or thrombosis. We conducted finite element analysis (FEA) of coronary stent to predict fatigue failure under cyclic bend loading. Methods: Matched with manufacturing steps of coronary stent, a stent with a diameter, length and thickness of 1.8mm, 18mm, and 81μm was modeled. The stent was crimped on a balloon by applying enforced displacement until the outer diameter became 1.0mm. Next, the stent was deployed into a bended coronary artery model with an angulation of 125 degrees. Then, cyclic bend from 125 degrees to 105 degrees was applied to the stent. Results and Discussions: FEA of alternating stress, mean stress, and fatigue safety factor showed that there were nodes in unsafe zone in Goodman diagram. FEA suggested a threshold of cyclic bend angulation of the coronary stent on fatigue failure. Fracture locations observed in the cyclic-bend accelerated durability tests was consistent with the nodes of the stents which exceeded fatigue safety factor of 1. Conclusion: The comparison between FEA and accelerated durability tests indicated that FEA was useful to predict fatigue failure of the coronary stent.layer and gas layer.
Content from these authors
© 2015 Japanese Society for Medical and Biological Engineering
Previous article Next article
feedback
Top