臨床応用を想定した直接給電FESの体内刺激装置実験に関する基礎的検討
倉田 絵莉宮原 敏佐藤 文博松木 英敏加藤 健治鈴木 迪諒澤田 真寛西村 幸男
著者情報
ジャーナル フリー

2015 年 53 巻 Supplement 号 p. S183_01

詳細
抄録

Functional electrical stimulation (FES) offers some mobility to individual with motor weakness. We propose the direct power feeding method for FES with mounted coils and miniature implanted electrodes. The mounted coils transmit power for driving the implants and communicate with them contactless via electromagnetic induction. These electrodes can stimulate muscle or nervous system locally. Furthermore, this method has a low risk of infection. Our group has developed the direct power feeding FES system, but not yet evaluated the system in animals. In this study, to document feasibility of this system to macaque monkeys, we investigated evoked movement of an upper extremity. In the experiment, the internal devices were not implanted and their output was connected to the connector extended from the electrodes implanted chronically in the muscle to the outside of the body. In the experiment of single channel stimulation, the desired response has been successfully observed on multiple muscles. Additionally, dual channel stimulation has successfully triggered multi-joint movements. Therefore, it has been confirmed that the direct feeding FES system can function successfully in vivo.