Transactions of Japanese Society for Medical and Biological Engineering
Online ISSN : 1881-4379
Print ISSN : 1347-443X
ISSN-L : 1347-443X
Proceedings
A fabrication of a combined treatment system of cryosurgery and hyperthermia treatment with a refrigerant circulation
DAISHI TAKAHASHIYOUSUKE TAGUCHITAKENORI AIDA
Author information
JOURNAL FREE ACCESS

2015 Volume 53 Issue Supplement Pages S208_03

Details
Abstract
The combination treatment system of cryosurgery and hyperthermia treatment was fabricated. The system consists of Peltier device, a power supply, a coolant chamber, Stirling cooler and a peristaltic pump. In the system fabrication, an aluminum box was mechanically worked and was attached to Stirling cooler and two pipes with a thermally-conductive bond. In the fabrication of the surgical probe, two pipes and Peltier device were respectively attached to two circular plates. The surgical probe was produced by fixing these plates to PVC pipe. Finally, the coolant chamber and the probe were connected utilizing the cold-resistant tube used at a peristaltic pump. Thus, the combination treatment system was fabricated. The system performance was evaluated using ethanol as a coolant in no-load running condition. During the test, a power was supplied to Peltier device and temperatures were measured. In the results of temperature measurement, the probe temperature reached to -45 degrees C and repeated the process of the cooling and the heating, alternately and continuously. This work was partly supported by JSPS Grants-in-Aid for Young Scientists (B) number 26820065.
Content from these authors
© 2015 Japanese Society for Medical and Biological Engineering
Previous article Next article
feedback
Top