Transactions of Japanese Society for Medical and Biological Engineering
Online ISSN : 1881-4379
Print ISSN : 1347-443X
ISSN-L : 1347-443X
Proceedings
A basic study on three-dimensional region-setting CT: development and evaluation of an experimental system
FUMIO HASHIMOTOATSUSHI TERAMOTOYASUKI ASADASHOICHI SUZUKIHIROSHI FUJITA
Author information
JOURNAL FREE ACCESS

2015 Volume 53 Issue Supplement Pages S211_02

Details
Abstract
Recently, region-setting CT has been studied as an ROI imaging method. This technique can obtain ROI image and strongly reduce the exposure dose by limiting the irradiation field. In this study, we developed three-dimensional region-setting CT and performed basic evaluations. The 6-channel collimators were controlled independently; it truncates projection data. Captured data were corrected using the mean pixel value of the valid data. The ROI image was reconstructed using the FDK algorithm. In the experiments, the image quality, shape reproducibility and exposure dose distribution were evaluated. These results indicate that proposed system may be useful for dose reduction in CT examination.
Content from these authors
© 2015 Japanese Society for Medical and Biological Engineering
Previous article Next article
feedback