3次元領域設定型CTに関する基礎的検討~実験装置の製作と評価~
橋本 二三生寺本 篤司浅田 恭生鈴木 昇一藤田 広志
著者情報
ジャーナル フリー

2015 年 53 巻 Supplement 号 p. S211_02

詳細
抄録
Recently, region-setting CT has been studied as an ROI imaging method. This technique can obtain ROI image and strongly reduce the exposure dose by limiting the irradiation field. In this study, we developed three-dimensional region-setting CT and performed basic evaluations. The 6-channel collimators were controlled independently; it truncates projection data. Captured data were corrected using the mean pixel value of the valid data. The ROI image was reconstructed using the FDK algorithm. In the experiments, the image quality, shape reproducibility and exposure dose distribution were evaluated. These results indicate that proposed system may be useful for dose reduction in CT examination.