Transactions of Japanese Society for Medical and Biological Engineering
Online ISSN : 1881-4379
Print ISSN : 1347-443X
ISSN-L : 1347-443X
Proceedings
Development of Capsuleless and Miniaturized Sweating Rate Meter Packaging Functional Elements
Takahiro KonoNoriko TsuruokaKeiji SuzukiTadao MatsunagaYoichi Haga
Author information
JOURNAL FREE ACCESS

2015 Volume 53 Issue Supplement Pages S486-S487

Details
Abstract
The purpose of this study is to develop a wearable sized sweating rate meter. It is known that thermal sweating rate reflects deep body temperature and dehydration. Therefore, continuous sweat measurement with a small sudorometer is expected to contribute to the prevention of heat stroke. The distinctive features of developed device are wearable size, eliminated capsules and packaged functional elements such as an air pump and two humidity sensors on it. Sweating rate is determined by the conventional method that uses the humidity difference between ambient air and the air moistened by sweat. The fabricated device has functional elements and flow channel on flexible substrate. It was evaluated that the device could quantify the sweating rate within precision of 15 percent by measuring water in place of sweat. Also, increases in thermal sweating rate on forearm after passive heat stress were successfully measured with the developed device in all subjects.
Content from these authors
© 2015 Japanese Society for Medical and Biological Engineering
Previous article Next article
feedback