The impact of dispersive electrode types on generation of low frequency noise
SHOTO YAMADARei EnatsuKYOSUKE YAMAMOTOKEITA OOMURARIKA KATOHIDEAKI TAMURATAKEO HASEGAWASHUICHI HASHIMOTOKANAE HASHIMOTOSHINYA CHIHARATAKAO MUROHASHI
Author information
JOURNAL FREE ACCESS

2018 Volume Annual56 Issue Abstract Pages S98

Details
Abstract

[Object] We evaluated the low frequency noise generated by two types of dispersive electrodes.[Method] Eight healthy subjects were enrolled in this study. The dispersive electrodes were set to the femur using two types of electrodes: conductivity type and capacitive type. The recording electrodes were placed on the inner vastus medialis (VM) and the rectus femoris muscle (RF). The low frequency noises (1-200 Hz) were compared between two types of dispersive electrodes.[Result] With respect to noises in VM, the mean volumes with capacitive type and conductivity type were 53.9± 3.16 dB and 48.0±1.03 dB, respectively. The mean RF noises with capacitive and conductivity type were 63.2± 1.73 dB, 56.5±1.51 dB, respectively. The noises of the capacitive type were significantly larger than the conductivity type.[Discussion] Our results indicate that the capacitive type of dispersive electrodes more affect IOM recordings than the conductivity type.