Transactions of Japanese Society for Medical and Biological Engineering
Online ISSN : 1881-4379
Print ISSN : 1347-443X
ISSN-L : 1347-443X
 
Six Degree Polished Titanium Taper & CoCr Head Had No Adverse Local Tissue Reaction in Total Hip Arthroplasty: 15-year Results
K HirakawaS TakayanagiN NakuraA SaitoS Ochiai
Author information
JOURNAL FREE ACCESS

2018 Volume Annual56 Issue Proc Pages 36-37

Details
Abstract

Several investigators reported that pseudotumor or adverse local tissue reaction (ALTR) occurred after total hip arthroplasty. Our study indicated that 6 degree polished titanium taper with smaller size or cobalt chromium (CoCr) femoral head had no ALTR compared to 12/14 taper rough surface finished taper junction with large diameter femoral head. These phenomena should be analyzed with cobalt chromium alloy trunnion taper fretting or crevice corrosion with titanium alloy femoral stem and CoCr head.

Content from these authors
© 2018 Japanese Society for Medical and Biological Engineering