Transactions of Japanese Society for Medical and Biological Engineering
Online ISSN : 1881-4379
Print ISSN : 1347-443X
ISSN-L : 1347-443X
Development of a Fracture-sign-detectable Retractor for Neurosurgery
RYOICHI MOCHIZUKIShingo TanakaYasuo SasagawaMitsutoshi NakadaTetsuyou Watanabe
Author information
JOURNAL FREE ACCESS

2021 Volume Annual59 Issue Proc Pages 743-745

Details
Abstract

Retractor is a medical device to expand the surgical field by compressing the tissue. However, there is a risk of serious accidents due to overloading in neurosurgery. In this study, we developed a retractor that detects signs of tissue fracture and automatically stop the compressing. We developed a retractor that can measure the pressure exerted on a tissue by measuring the pressure of incompressible fluid filled in a rubber bag attached on the retractor. The compression test of gelatin was conducted with a constant compression speed. The pressure behavior initially increases nonlinearly with the deformation of the target, then becomes linear, and then becomes nonlinear with crack initiation. We developed a system that automatically detects the point where the system starts becoming linear. We also developed a smaller retractor to expand the range of application. The effectiveness of the system was confirmed through verification experiments using tofu and sheep brains.

Content from these authors
© 2021 Japanese Society for Medical and Biological Engineering
Previous article Next article
feedback
Top