医用電子と生体工学
Online ISSN : 2185-5498
Print ISSN : 0021-3292
ISSN-L : 0021-3292
皮膚インピーダンスの等価回路の構成法
山本 尚武山本 辰馬
著者情報
ジャーナル フリー

1976 年 14 巻 2 号 p. 129-132

詳細
抄録
This paper explains the construction of equivalent circuit with fixed elements for skin impedance which satisfy the Cole-Cole circular arc law.
The Cole-Cole dispersion system having continuously distributed relaxation times can be simulated by the system of some discrete relaxation times, and approximate equation is assumed as follows :
P (ω) =Z (ω) -Z/Z0-Z=1/1+ (jωτm) βni=1Ri/jωCiRi/ni=1Ri
where Z (ω) is skin impedance and the suffixes 0 and co refer to zero and infinite angular frequency.βandτmare parameters of the Cole-Cole arc. To determine the optimum coefficients Ri Ci of thisequation some conditions and evaluation function are provided. Ri, Ci< are frequency-independent fixed elements. Then, optimum coefficients were obtained for β=0.5-0.9 for the cases n=3, 5, 7. The equivalent circuit, constructed by n series arrangements of parallel resistance Ri and capacitance Ci combination and series resistance Z, simulates in good agreement skin impedance over the wide frequency range. It is useful to the simulator of skin impedance in the case, for example, of signal distortion caused by skin impedance appearing : in, electrocardiography.
著者関連情報