医用電子と生体工学
Online ISSN : 2185-5498
Print ISSN : 0021-3292
ISSN-L : 0021-3292
適応形相関フィルタによるP波検出に及ぼすデータ圧縮の影響について
村上 啓二田中 弘富増山 善明吉川 昭
著者情報
ジャーナル フリー

1982 年 20 巻 2 号 p. 113-116

詳細
抄録

Effects of data compression, to save the processing time in the P-wave recognition program with the adaptive correlating filter (ACF), on the performance of the ACF has been studied.All QRS or QRST regions were eliminated from the record to compress ECG data.
Two indices were adopted as criteria for evaluating the effects. One was derived on the basis of the performance characteristics of the ACF, while the other was the mean error probability in detecting P-waves by using the ACF.
In result, it was shown that the method for data compression could make up for the loss of the P-wave information caused by the data reduction. So the data reduction proposed here makes data compression feasible with little deterioration of the ACF performance.

著者関連情報
© 日本生体医工学会
前の記事 次の記事