高速液晶を用いたフリッカ値測定装置の試作
保坂 良資池田 研二斎藤 正男西野 治
著者情報
ジャーナル フリー

1983 年 21 巻 6 号 p. 472-475

詳細
抄録
A new device to measure the critical-fusion frequency (CFF) has been developed. In conventional methods, a light-source is set in a proper background and is made on-and-off. By such a type of equipment, measurement during work is difficult, since the examinee have to look into a dark box. The equipment is large in size, because the equipment includes some mechanical moving parts. In the proposed equipment, natural light is used as a source, which is modulated by a liquid crystal. This new equipment is small in size and makes possible the measurement of CFF during work, since a flickering spot appears in the open visual field of the examinee. A liquid crystal has been developed for this purpose with the response times of 1ms from clear to dark state and 8ms from dark to clear state, which is enough to measure ordinary CFF ranging up to 50Hz.