医用電子と生体工学
Online ISSN : 2185-5498
Print ISSN : 0021-3292
ISSN-L : 0021-3292
ヒト錐体視物質濃度変化の他覚的測定装置の製作
王 景雪斎藤 昭三宅 養三神保 孝志梅野 正義
著者情報
ジャーナル フリー

1991 年 29 巻 4 号 p. 249-253

詳細
抄録
This paper describes a fundus reflection densitometer and its application to measurement of the foveal cone pigment regeneration of healthy subjects. A 500W xenon lamp provides bleaching, measuring and reference light beams. A PC9801 personal computer is used for control shutters of three beams, on-line calculation and display of cone pigment regeneration. When the bleaching light is on, the fundus can be viewed. The measuring field is fixed at 3.5deg. Using it, we have measured the two-way density and regeneration time of 5 healthy subjects (3 male and 2 female), with clear optical media and good visual acuity, ranging in age from 27 to 28 years. The two-way densities and regeneration times of 5 subjects show mean values of 0.426±0.038 SD and 152.5±21.1 SD s, respectively.
著者関連情報
© 日本生体医工学会
前の記事 次の記事
feedback
Top