Magnetic Resonance Imaging-Microdialysis Simultaneous Measuring System : Neurotransmitter Levels in a Rat Brain during Magnetic Resonance Imaging
Koji HYODOYumiko WATANABEKazuhiro HOMMA
著者情報
ジャーナル フリー

1999 年 42 巻 3 号 p. 784-789

詳細
抄録
A new simultaneous measuring system using Magnetic Resonance Imaging (MRI) and microdialysis was developed. Because MRI and microdialysis have complementary measuring advantages, the combination of the two methods can provide less invasiveness, high detection sensitivity and multimodality for in vivo biomedical measurement. Using the MRI-Microdialysis simultaneous measuring system, preliminary experiments were performed to examine the rapid effect which MRI has on rat brain metabolism and to analyze the acute physiological response of the brain to oxidative stress. The changes in 5-hydroxyindoleacetic acid (5HIAA), noradrenaline levels in the striatum and 5HIAA, L-aspartic acid and L-asparagine levels in the hypothalamus area were not significant when compared to normal changes in the absence of the MRI sequence. There were no significant changes in 5-HIAA and glutamine levels during the oxidative stress;however, the intensity of the MR image was increased. The increase suggested that a decrease in paramagnetic molecules occurred during the oxidative stress period in the brain.
著者関連情報
© The Japan Society of Mechanical Engineers
前の記事 次の記事
feedback
Top