The Proceedings of the International Conference on Motion and Vibration Control
Online ISSN : 2424-2977
6.2
会議情報
ADAPTIVE VIBRATION CONTROL IN AN IMPLANTABLE BLOOD PUMP ON A BEARINGLESS SLICE MOTOR
Christian Huettner
著者情報
会議録・要旨集 フリー

p. 684-689

詳細
抄録
Implantable left ventricular assist devices are powered by batteries. Their limited capacity has to be used as efficiently as possible. At the example of a MAGLEV centrifugal LVAD it was demonstrated that a drastic reduction of power consumption (50% in the bearing) could be achieved by vibration control. A vibration controller design for the non-linear plant is discussed, simplified and extended to four harmonics. Anew phase shift strategy is presented to eliminate time-consuming matrix operations. Based on this simplification the implemented algorithm applied on the bearingless slice motor allocates little memory and requires short computation time. Additionally, the decreased control currents lead to higher control margins of the power amplifiers and therefore to improved robustness.
著者関連情報
© 2002 The Japan Society of Mechanical Engineers
前の記事 次の記事
feedback
Top