In Magnetic Resonance Imaging (MRI) apparatus, pulse current is energized to the gradient coils in a strong static magnetic field generated by the static magnetic poles. Since electromagnetic force (i.e. Lorentz force) is generated in the gradient coils, the MRI magnet system vibrates. On the other hand, vibration of the MRI magnet system is affected by electromagnetic force caused by static magnetic poles vibration. As the vibration of MRI magnet system causes magnetic field disturbance (so-called "error magnetic field") and affect image quality, it is important to evaluate them in the design process. In this study, a varying magnetic field evaluation method for MRI magnet system was developed. Vibration and electromagnetic force is considered in the weak coupling formation using the Modal Magnetic Dumping (MMD) method. In the eddy current analysis by vibration, the displacement was considered in the magnetic field changes in the finite elements. Error magnetic field caused by equipment vibration was obtained by superposition of the static magnetic field fluctuation and the eddy current magnetic field. The open type MRI magnet was evaluated by the proposed methodology. As a result, vibration of static magnet poles were suppressed by magnetic dumping at 50Hz or less and eddy current magnetic field was dominant at 50Hz or more.