255 手保持型耳小骨可動性診断装置の開発
小池 卓二和田 仁湯浅 有湯浅 涼川瀬 哲明小林 俊光
著者情報
会議録・要旨集 フリー

p. 129-130

詳細
抄録
It is important to evaluate stapes mobility in tympanoplasty surgery, because it affects the prognosis of improvement of the hearing level. In this study, a new hand-guided apparatus was developed by assembling a small-sized force sensor and a hydraulic micromanipulator. Then, the mobility of stapes in humans was measured. In addition, the nobilities of incus and malleus in guinea pigs are also measured under the condition that the stapes was artificially fixed with glue being applied to the annular ligament to simulate otosclerosis. The load-displacement curve obtained from the normal stapes was different from that obtained from the fixed stapes. This result suggests that a surgeon can quantitatively analyze stapes mobility with this apparatus during surgery.
著者関連情報
© 2002 一般社団法人日本機械学会