205 インスリンポンプ用熱式超微量流量計の基礎研究
西山 尚小西 義昭藤原 真人直野 義昭角田 直人山田 幸生
著者情報
会議録・要旨集 フリー

p. 17-18

詳細
抄録
The flow rate of insulin from insulin pumps for diabetic patients is less than 3.0 μl/min, and the average flow velocity in a tube with a diameter of 0.4 mm becomes less than 0.4 mm/s resulting in the maximum Reynolds number of about 0.2. A thermo-trace type flow meter is employed for a low meter in this very small range of flow rate and velocity. A theoretical analysis and preliminary experiment were conducted for the flow meter. It was found that the influence of heat conduction in the direction of the tube axis must be included which is usually ignored in the boundary flow analyses.