517 動圧浮上遠心式血液ポンプの血液適合性の検討
山根 隆志加藤 孝久丸山 修西田 正浩大久保 剛佐野 岳志宮本 祐介
著者情報
会議録・要旨集 フリー

p. 79-80

詳細
抄録
A hydrodynamic levitation centrifugal pump is being developed for a long-term implantable artificial heart. The hemocompatibility was investigated with numerical analysis and has been verified with animal blood tests. The hemolytic property was improved by expanding the bearing gap and by adjusting the pressure balance and the antithrombogenicity has been improved by adding deep grooves to the hydrodynamic bearings to increase the flow rate through the bearings.