522 セルソーターの高速化に関する研究
宮崎 康次飯田 剛士堀尾 寛塚本 寛
著者情報
会議録・要旨集 フリー

p. 89-90

詳細
抄録
An experimental study on droplet generation for the rapid detection of cells by using cell sorters was conducted. Cell sorters create micro-droplets by controlling waves at surface of a micro-jet which consists of samples and sheath fluid. In this paper the optimum wave number was observed by microscope The ratio of sample flow rates to sheath flow rates and frequency of the wave are also controlled in order to investigate those effects on droplet generation. In addition to the experimental studies linear instability analysis for droplet generation was carried out to obtain optimum parameters to maximize the rate of droplet generation. Analytical solutions agree well with the experimental results of droplet generation.
著者関連情報
© 2003 一般社団法人日本機械学会