年次大会講演論文集
Online ISSN : 2433-1325
セッションID: 0911
会議情報
0911 HMS用自動血管位置探索デバイスの開発(J07-3 バイオ・医療と設計(3),J07 バイオ・医療と設計)
神谷 宏文上野谷 敏之仲町 英治上辻 靖智
著者情報
会議録・要旨集 フリー

詳細
抄録
In this study, an automatic blood vessel searching system (BVSS) is newly developed, which is built in the health monitoring system (HMS) and the drug delivery system (DDS) to evaluate the blood sugar level and inject the insulin for the diabetic patients. Main subjects of this BVSS development are 1) a transmittance photo imaging of the finger by using the LED light as a near-infrared light source with peak wavelength of 870nm, and 2) an image processing to detect the location of the center of blood vessel. To estimate the depth of blood vessel, the lens focal method was applied to the BVSS. To confirm the availability of the system, we used blood vessel phantoms that consist of a cylindrical blood vessel (metal rod) and skin tissue (silicone rubber), and confirmed the artificial blood vessel was 1.5mm in diameter. The phantoms were embedded in the skin under 1.7mm in depth. Estimated depths using our image processing show good agreements with the given depths.
著者関連情報
© 2007 一般社団法人日本機械学会