The proceedings of the JSME annual meeting
Online ISSN : 2433-1325
2007.6
Session ID : 0913
Conference information
0913 Development of Automated blood extraction device with microneedle for HMS/DDS
Kaichirou KAWANAKAToshiyuki UENOYAEiji NAKAMACHIYasuyomo UETSUJI
Author information
CONFERENCE PROCEEDINGS RESTRICTED ACCESS

Details
Abstract

The purpose of this research project is to develop a compact and wearable Health Monitoring System(HMS) device which acts automatically 1) puncturing a blood vessel, 2) extracting human blood and 3) measuring blood glucose. In this study a permanent magnet driven blood extraction pump is newly developed which has a painless microneedle with inner diameter less than 100μm. The pump consists of a) puncture part, and b) syringe part which is inserted a permanent magnet with magnetic fluid for sealing action and lubrication. The result of suctioning experiment using the pump, it succeeded in extracting 7.9μl of water within 6 seconds.