1P1-N02 Magnetically driven mechanisms for cytology brush built into capsule endoscope(Medical Robotics and Mechatronics (1))
Maki YamashitaShintaro YamadaTakashi Honda
Author information
CONFERENCE PROCEEDINGS FREE ACCESS

Details
Abstract
In this paper, we propose two types of magnetic microactuators for a cytology brush built into a capsule endoscope. One is a moving-magnet linear actuator, in which rotational oscillation of a permanent magnet is converted into linear movement through friction. The other is a rotating-magnet actuator with a flexible joint. Both actuators feature wireless remote drive by external magnetic field. As a result of the operation test using raw meat, the brush was successfully driven. In addition, we have applied a new function of angular positioning to the capsule by rotating magnetic field.